5-8
Chapter 5 Initial Settings
Alarm Related Setup
CAUTION
When "Suspend Arrhy. Analysis during Noise Interference" is set to [ON], and the
suspended duration continues for more than 30 seconds, "Cannot analyze" message
will generate.
9
Set the "Lower Limit for Alarm Volume".
WARNING
Changing the setting for "Alarm System" will also change the alarm volume and tone
setting. As the "Lower Limit for Alarm Volume" may also change, make sure to check the
volume and tone on the "Tone/Volume" setup menu.
1
Press the key for "Lower Limit for Alarm Volume".
The "Lower Limit for Alarm Volume" window will be displayed.
2
Set the lower limit of alarm volume for "Vital Alarm", "Ventilator Alarm", "Status Alarm".
The lower limit of adjustable alarm volume range on the "Tone/Volume" setup menu will be set. The
lower limit level can be set according to the alarm level priority, Urgent>Caution>Status.
[Test]: The test sound will be generated with the set volume.
10
Set the alarm indicator operation.
NOTE
The alarm indicator setting is to be performed for each alarm level.
REFERENCE
The alarm indicator flashing pattern can be set according to the alarm level. The patient's
condition can be checked from far distance by the difference of flashing pattern.
1
Press the key for the level to set the flash pattern.
The pattern setup window will be displayed.
Alarm Indicator Flashing Pattern
Pattern
Flash Pattern
Pattern A
(Red, Red, Red), (xxx), (Red, Red, Red), (xxx), (Red, Red, Red)
Pattern B
(Red, Orange, Red), (xxx), (Red, Orange, Red), (xxx), (Red, Orange, Red)
Pattern C
(Red, Green, Red), (xxx), (Red, Green, Red), (xxx), (Red, Green, Red)
Pattern D
(x , Red, x), (xxx), (x, Red, x), (xxx), (x, Red, x)
Pattern E
(x, Orange, x), (xxx), (x, Orange, x), (xxx), (x, Orange, x)
Pattern F
(x, Blue, x), (x, Blue, x), (x, Blue, x), (x, Blue, x), (x, Blue, x),
Pattern G
(Red, Red, x), (xxx), (x, Red, Red), (xxx), (Red, Red, x)
Pattern H
(Red, Orange, x), (xxx), (x, Orange, Red), (xxx), (Red, Orange, x)
Pattern I
(Red, Green, x), (xxx), (x, Green, Red), (xxx), (Red, Green, x)
Pattern J
(Red, x, x), (xxx), (x, Red, x), (xxx), (x, x, Red)
* (xxx) indicates that the alarm indicator is not lit.
Summary of Contents for 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...6 Contents ...
Page 10: ...Contents ...
Page 16: ...Contents ...
Page 46: ...xxx Safety Electromagnetic Compatibility ...
Page 48: ...Chapter 1 Installation of the Unit Contents ...
Page 70: ...Chapter 2 Network System Construction Contents ...
Page 82: ...2 12 Chapter 2 Network System Construction Wireless Network ...
Page 84: ...Chapter 3 Using the CF card Contents ...
Page 90: ...3 6 Chapter 3 Using the CF card Formatting the SD Card ...
Page 92: ...Chapter 4 Connection to the External Devices Contents ...
Page 124: ...4 32 Chapter 4 Connection to the External Devices Connection with the Laser Printer ...
Page 126: ...Chapter 5 Initial Settings Contents ...
Page 158: ...5 32 Chapter 5 Initial Settings User Mode Registration ...
Page 160: ...Chapter 6 Setup Item Default Value Contents ...
Page 200: ...Chapter 7 Replacement Parts Contents ...
Page 204: ...Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 212: ...8 8 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 214: ...Chapter 9 Maintenance Check Contents ...
Page 228: ...Index 2 Index ...
Page 229: ......
Page 230: ......