8-4
Chapter 8 Cleaning/Disinfecting/Storing
Cleaning the Equipment and Sensors
Fan Filter
The main unit incorporates a fan. Check it periodically and when it gets dusty, clean the filter. The filter can be
detached. Make sure to attach it properly after cleaning.
CAUTION
After cleaning the filter with neutral detergent, dry it completely before attaching. If the
moisture is remained on the filter, it may damage the equipment.
After cleaning the filter, make sure to reattach it to the equipment. If the equipment is used
without the filter, it may damage the equipment.
NIBP Cuff and Air Hose
To clean the cuff shell, remove the bladder and wash it using neutral detergent. Make sure that the cuff is dry before
placing back the bladder inside.
For procedure to clean/disinfect the reusable cuff, refer to the instruction manual for the respective cuff.
Do not reuse/resterilize the disposable cuff.
CAUTION
After washing, ensure the size indication on the bladder and cuff shell match. Make sure that
the cuff hose is threaded through one of the hose openings in the cuff.
BP Transducer
Disinfect the blood pressure transducers according to the manufacturer’s guidelines.
Do not reuse / re-sterilize the disposable type transducers.
SpO
2
Sensor
Disinfect the SpO
2
sensor according to the manufacturer's guidelines. Do not reuse/resterilize the disposable SpO
2
sensor.
Nellcor Sensor
Do not soak the sensor in water or antiseptic solution.
Wipe the DURASENSOR with disinfectant such as 70% alcohol. Do not sterilize by irradiation, steam, or
ethylene oxide.
OxiMax is a single-patient use type sensor. Do not reuse or resterilize.
Masimo Sensor
Do not soak the sensor or patient cable in water or antiseptic solution.
(Sensors and connectors are not water-proof.)
Do not sterilize the sensors and cables by irradiation, steam, or ethylene oxide.
The Masimo disposable sensor can be reused on the same patient if the light emitting and receiving part is
clean, and if it is still adhesive to the skin.
For most of the sensors, the adhesiveness will return by cleaning the sensor with alcohol and completely drying
Summary of Contents for 8000 Series
Page 1: ......
Page 2: ......
Page 8: ...6 Contents ...
Page 10: ...Contents ...
Page 16: ...Contents ...
Page 46: ...xxx Safety Electromagnetic Compatibility ...
Page 48: ...Chapter 1 Installation of the Unit Contents ...
Page 70: ...Chapter 2 Network System Construction Contents ...
Page 82: ...2 12 Chapter 2 Network System Construction Wireless Network ...
Page 84: ...Chapter 3 Using the CF card Contents ...
Page 90: ...3 6 Chapter 3 Using the CF card Formatting the SD Card ...
Page 92: ...Chapter 4 Connection to the External Devices Contents ...
Page 124: ...4 32 Chapter 4 Connection to the External Devices Connection with the Laser Printer ...
Page 126: ...Chapter 5 Initial Settings Contents ...
Page 158: ...5 32 Chapter 5 Initial Settings User Mode Registration ...
Page 160: ...Chapter 6 Setup Item Default Value Contents ...
Page 200: ...Chapter 7 Replacement Parts Contents ...
Page 204: ...Chapter 8 Cleaning Disinfecting Storing Contents ...
Page 212: ...8 8 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Page 214: ...Chapter 9 Maintenance Check Contents ...
Page 228: ...Index 2 Index ...
Page 229: ......
Page 230: ......