Splice Menu
(4) The calibration result is displayed after completion.
“Test Finished” message
Arc power and splicing
position calibration were
successfully completed. Press
to exit.
“Test Again” message
The arc power and splicing
position calibration were
completed but further
calibration is strongly
recommended, as the change
from the previous arc
calibration is too large.
Press to perform arc
calibration, or to exit even
though arc calibration is not
complete.
• In some cases, multiple iterations of
arc calibration are needed until the
calibration process is successfully
completed and the "Test Finished"
message is displayed. Arc
calibration can be considered almost
complete if multiple iterations are
completed without receiving the
message.
• A specific numeric threshold can be
set so that "Test Finished" message
is displayed after a specific number
of arc calibrations are performed.
See [Other Option] (Page 55)
for
detail.
Arc calibration
Finish
Arc calibration
Select [Arc Calibration]
in the [Splice Menu]
Arc calibration
Test Again”