71
Questions and Troubleshooting
(3) Method to set parameters of Splice or Heater mode from a PC
• To set or change the parameters, contact your nearest sales agency.
(4) Forgot password
• Contact your nearest sales agency.
5. Other Functions
(1) Method to execute [Dust Check] automatically after turning power on
• Refer to [Power On Option] (Page 54).
(2) Method to hide messages on [READY] screen
• Change the fiber image from X/Y view by pressing .
(3) Too many repetitions until “Test Finished” indicated in [Arc Calibration]
• The splicer needs to repeat the arc calibration after replacing the electrodes or when the
environmental conditions change drastically. (Page 55). The number of arc calibrations
can be set to a specific amount. When the splicer completes the set amount of
calibrations, it indicates “Test Finished”. However, this does not mean it is completely
calibrated.
(4) “Test Finished” is never indicated after many repetitions in [Arc Calibration]
• Execute [Stabilizing Electrodes] function in [Maintenance Menu]. If the splicer still
does not indicate “Test Finished”, replace the electrodes. Refer to [Replace Electrode]
section.
(5) No arc power change after [Arc Calibration]
• An internal factor is calibrated and adjusted for the specific arc power selected. The
displayed arc power in each splice mode does not change.
(6) Method to input different comments after each splice in splice results data
• Refer to [Splicing procedure] (Page 20).
(7) Method to input the same comments automatically in splice results data
• Refer to [Splicing procedure] (Page 20).
(8) Method to cancel storing splice results in internal memory
• Refer to [Splicing procedure] (Page 20).
(9) Method to download splice results from splicer to PC
• Contact your nearest sales agency.