67
Error Message List
Error
Message
Reason
Solution
Arc
Discharge
Delayed
Arc Discharge is
delayed
• Check that the electrodes are in proper
position.
• Execute the [Stabilize Electrode] function.
• Replace the electrodes.
Inadequate arc power
Calibrate the arc power with the [Arc
Calibration] function.
Prefuse power or time
is set too high
If using the standard mode, adjust or initialize
[Prefuse Power] or [Prefuse Time] settings. For
AUTO splice mode, the prefuse is fixed and
cannot be adjusted.
Thin Fiber
Insufficient [Overlap]
setting
Fat Fiber
Too much [Overlap]
setting
If using the standard mode, adjust or initialize
[Overlap] setting. For AUTO splice mode, the
overlap is fixed and cannot be adjusted.
Too Tapering
Fiber
Too much fiber taper.
If using the taper splice function, the error
message may appear meaning the fiber is
tapered too much.
Bad fiber end-face.
Check the condition of fiber cleaver. When the
blade is worn, rotate the blade.
Bubble
Prefuse power or time
is set too low
If using a standard mode, adjust or initialize
[Prefuse Power] or [Prefuse Time] settings. For
AUTO splice mode, the prefuse is fixed and
cannot be adjusted.
There is dust
after
executing
Dust Check
function
Dirt or dust exists in
optical path.
• Clean the objective lenses by referring page to
28.
• When the above processes cannot remove the
dirt or dust, consult your nearest sales agency.
Heater Oven
Trouble
Heater does not heat.
Consult your nearest sales agency.
Temperature
Sensor NG
Temperature Sensor
may be damaged.
Consult your nearest sales agency.