When test pattern film is output after a while, make sure that image has been printed
correctly.
[Fig. 4.18] Standard Test Pattern (SMPTE) - Example: For an output image
30%
40%
50%
50%
60%
20%
10%
0%
70%
80%
90%
100%
0/
5%
95/
100%
SMPTE TEST PATTERN
REV. 1 0 / 6 / 8 3 © 1 9 8 3
9 / 6 / 8 4
When the Service Setting has been performed, a user-defined test pattern can be selected.
From the Test Pattern screen, select the “
User Settable
”.
Then, follow the same operation procedures as for the SMPTE pattern.
[Fig. 4.19.1]
When the Service Setting has been performed, a test pattern to be used for confirmation of
the resolution can be selected.
From the Test Pattern screen, select the “
Spatial Resolution
”.
Printout will then start after Dmax is selected.
[Fig. 4.19.2]
Select.
Select.
Summary of Contents for Drypix 5000
Page 7: ...1 1 006 240 20 2004 03 1 Chapter 1 Introduction ...
Page 17: ...2 1 006 240 20 2004 03 2 Chapter 2 System Overview ...
Page 23: ...3 1 006 240 20 2004 03 3 Chapter 3 Basic Operation ...
Page 40: ...3 18 006 240 20 2004 03 Basic Operation ...
Page 41: ...4 1 006 240 20 2004 03 4 Chapter 4 Utility Operation ...
Page 87: ...5 1 006 240 20 2004 03 5 Chapter 5 Troubleshooting ...
Page 125: ...6 1 006 240 20 2004 03 6 Chapter 6 Regular Care and Maintenance ...
Page 130: ...6 6 006 240 60 2007 03 Regular Care and Maintenance ...
Page 131: ...7 1 006 240 20 2004 03 7 Chapter 7 Specifications ...
Page 136: ...7 6 006 240 20 2004 03 Specifications ...
Page 138: ...Maintenance and Inspection 006 240 20 2004 03 ...
Page 141: ......
Page 142: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A ...