1-4
006-240-60 2007.03
Introduction
1.4 Operational Precautions
1.4.1 Laser Handling Precautions
DRYPIX 5000 Laser Unit Specifications
Class
3B
Medium
Semiconductor laser
Wavelength
660 nm
Maximum Output
50 mW (CW)
WARNING
The DRYPIX 5000 incorporates a Class 3B laser with a maximum output of
50mW.
To prevent exposure to its laser beams, observe the following precautions.
• Never remove the protective covers.
• When opening the outer covers for coping with film jams, be sure to observe
procedures set forth in the Operation Manual.
• Always contact a qualified service engineer immediately if you suspect there
is a malfunction.
CAUTION
Use of controls or adjustments or performance of procedures other than those
specified herein may result in hazardous radiation exposure.
1.4.2 Preventing Electric Shock
WARNING
The power supply to the DRYPIX 5000 is AC100/110/120V.
To avoid possible electrical shock, users should observe the following precau-
tions.
• Avoid installation sites where water may splash on the equipment.
• Make sure that the equipment is properly grounded to a protective earth lead
for indoor wiring.
• Make sure that all cable connections have been properly connected.
WARNING
Inside the DRYPIX 5000 are parts through which a high voltage is conducted,
putting the user at risk for electric shock. Never remove any covers other than
the eject and front covers.
1.4.3 Ventilation Precautions
CAUTION
Be sure to install this equipment in a well-ventilated environment.
Good ventilation must attend its use (at least 54m
3
/hour for one unit).
1.4.4 Handling Precautions
CAUTION
If the room temperature increases abruptly when the equipment is cold, dew
condensation may occur. Exercise care so that dew condensation does not
occur because it may cause the equipment to be damaged.
Summary of Contents for Drypix 5000
Page 7: ...1 1 006 240 20 2004 03 1 Chapter 1 Introduction ...
Page 17: ...2 1 006 240 20 2004 03 2 Chapter 2 System Overview ...
Page 23: ...3 1 006 240 20 2004 03 3 Chapter 3 Basic Operation ...
Page 40: ...3 18 006 240 20 2004 03 Basic Operation ...
Page 41: ...4 1 006 240 20 2004 03 4 Chapter 4 Utility Operation ...
Page 87: ...5 1 006 240 20 2004 03 5 Chapter 5 Troubleshooting ...
Page 125: ...6 1 006 240 20 2004 03 6 Chapter 6 Regular Care and Maintenance ...
Page 130: ...6 6 006 240 60 2007 03 Regular Care and Maintenance ...
Page 131: ...7 1 006 240 20 2004 03 7 Chapter 7 Specifications ...
Page 136: ...7 6 006 240 20 2004 03 Specifications ...
Page 138: ...Maintenance and Inspection 006 240 20 2004 03 ...
Page 141: ......
Page 142: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A ...