Basic Operation
1
2
3
4
5
6
7
006-240-30 2006.05
3-11
5-8
The display returns to the routine screen. This completes the film supply
procedure. Note, however, that if you have selected
By each film pack
for
AUTOMATED F.D.C
. in the Utility mode, the Auto F.D.C. (automatic film density
calibration) will be performed. After completion of the density calibration
procedure, the display returns to the routine screen.
For details of
AUTOMATED F.D.C.
, see
Chapter 4, Utility Operation
.
6
When an incorrect film pack was loaded or films of unspecified type were loaded
6-1
When an incorrect film pack was loaded:
When the film pack is loaded incorrectly (loaded up-side-down, one end of the film pack is
not outside the tray), the related tray will be unlocked. Then, the following screen will
appear.
If you want to stop loading films while the following screen
[Fig. 3.21]
is being
displayed on the operation panel, remove the film pack and close the tray.
button will appear on the operation panel.
Press the
button, and open and then close the tray again.
[Fig. 3.21]
Summary of Contents for Drypix 5000
Page 7: ...1 1 006 240 20 2004 03 1 Chapter 1 Introduction ...
Page 17: ...2 1 006 240 20 2004 03 2 Chapter 2 System Overview ...
Page 23: ...3 1 006 240 20 2004 03 3 Chapter 3 Basic Operation ...
Page 40: ...3 18 006 240 20 2004 03 Basic Operation ...
Page 41: ...4 1 006 240 20 2004 03 4 Chapter 4 Utility Operation ...
Page 87: ...5 1 006 240 20 2004 03 5 Chapter 5 Troubleshooting ...
Page 125: ...6 1 006 240 20 2004 03 6 Chapter 6 Regular Care and Maintenance ...
Page 130: ...6 6 006 240 60 2007 03 Regular Care and Maintenance ...
Page 131: ...7 1 006 240 20 2004 03 7 Chapter 7 Specifications ...
Page 136: ...7 6 006 240 20 2004 03 Specifications ...
Page 138: ...Maintenance and Inspection 006 240 20 2004 03 ...
Page 141: ......
Page 142: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A ...