5-31
006-240-30 2006.05
Troubleshooting
1
2
3
4
5
6
7
5.5.3 When a new film pack was loaded and the tray closed, but the positioning
was wrong:
[Fig. 5.36]
will appear.
Since the tray has not been locked, pull out the tray.
Place the film pack in the correct position and pull out the film end, and close the tray.
[Fig. 5.38]
●
Placed up-side down
●
Placed front side back
To suspend film replenishment, press the
button and open/close the tray
again.
5.5.4 After placing a new film pack and pulling out the film pack,
the tray was closed without cutting the bottom end of the film pack:
In this case, since the bottom end of the film pack is not cut out, the film pack cannot be
removed.
Press the
button to unlock.
Pull out the tray and cut the bottom of the film pack with a cutter, and push back the tray
in.
[Fig. 5.39]
If the film pack is harshly deformed, remove the pack and place a new one.
CAUTION
If the cutter’s blade protection is broken, the blade will be exposed posing then
danger to the user. Do not touch the unprotected blade.
Replace a dull cutter hard to cut the film pack, as necessary.
Cutter
Summary of Contents for Drypix 5000
Page 7: ...1 1 006 240 20 2004 03 1 Chapter 1 Introduction ...
Page 17: ...2 1 006 240 20 2004 03 2 Chapter 2 System Overview ...
Page 23: ...3 1 006 240 20 2004 03 3 Chapter 3 Basic Operation ...
Page 40: ...3 18 006 240 20 2004 03 Basic Operation ...
Page 41: ...4 1 006 240 20 2004 03 4 Chapter 4 Utility Operation ...
Page 87: ...5 1 006 240 20 2004 03 5 Chapter 5 Troubleshooting ...
Page 125: ...6 1 006 240 20 2004 03 6 Chapter 6 Regular Care and Maintenance ...
Page 130: ...6 6 006 240 60 2007 03 Regular Care and Maintenance ...
Page 131: ...7 1 006 240 20 2004 03 7 Chapter 7 Specifications ...
Page 136: ...7 6 006 240 20 2004 03 Specifications ...
Page 138: ...Maintenance and Inspection 006 240 20 2004 03 ...
Page 141: ......
Page 142: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A ...