6.7 Analog input
By providing the signals from external sensors to the control unit, you can reflect the changes in gas condi-
tions into the measurement result. The analyzer can accept two 4–20 mA DC inputs from a thermometer,
pressure gauge, flowmeter, oxygen analyzer, and/or moisture meter. Wire between each sensor and the ana-
log input terminal of the control unit in reference to “3.2 Wiring diagram”.
Set the parameters of an item with no sensor input to “fixed” and enter a value for each. For items whose
change is negligible, you can set the item to a fixed value.
Be sure to make settings for all the items. Otherwise, the analyzer cannot deliver accurate measurement.
Select “Input Signal” from the “Menu” screen, and press the
key.
6.7.1 Analog input setting: sensor input
The description here is an example when using a thermometer.
Connect the 4-20 mA DC signal from a thermometer to Ch1 or Ch2 of AI terminal.
If you make a setting but there is no signal input, the analyzer emits an “analog input error” alarm.
Select the “Gas Temp.” by the
key or the
key. Press the
key to move the cur-
sor to “Channel” setting.
(
)
ENT
Summary of Contents for ZSS-8 Series
Page 1: ......
Page 13: ...INZ TN8ZSS E 3...
Page 16: ...6 INZ TN8ZSS E 3 2 Wiring diagram Input output terminal...
Page 17: ...INZ TN8ZSS E 7...
Page 29: ...INZ TN8ZSS E 19 4 8 Piping system diagram Fig 4 9...
Page 42: ...32 INZ TN8ZSS E 5 2 Screen configuration...
Page 57: ...INZ TN8ZSS E 47 About Span Calibration...
Page 76: ...66 INZ TN8ZSS E...
Page 111: ...INZ TN8ZSS E 101 APPENDIX 2 CODE SYMBOLS...
Page 118: ...108 INZ TN8ZSS E DIMENSIONS IN MM...
Page 119: ......