4.7.2 How to use the angle adjustment unit 2
This section describes movement of the optical axis by operating the angle adjustment unit.
Example) When you adjust the fixing bolt and the anchor nut shown in the figure below, encircled with a
solid line
(1) If the flange A and the flange B are parallel, the laser beam points horizontal direction.
(2) If you screw up the bolt shown in the below figure, the place where the bolt is screwed up moves to-
wards the flange A, and the laser beam faces downward.
(3) If you loosen the fixing bolt as the figure below, the place where the bolt is loosen aparts from the
flange A, and the laser beam point upward.
・
In the same way, if you screw up the bolt and nut colored gray in the below figure, the light axis moves as
follows.