INZ-TN8ZSS-E
- 51 -
Point the cursor to [Cal Gas Value] with
and then press the
key to
reverse out the cell.
By using the
keys, enter the cal-
ibration gas concentration calculated in
6.2.3.2, and then click the
key.
Point the cursor to [Gas Temp] with
and then press the
key to reverse
out the cell.
Input the 25 [°C] with
and then
press .
Point the cursor to [Gas Press] with
and then press the
key to reverse
out the cell. Input the 0.00[kPa] with
and then press
.
ENT
ENT
ENT