- 50 -
INZ-TN8ZSS-E
6.2.3.3 Procedure of span calibration
In [Menu] screen, point the cursor to [span
calibration]
with
and
then
press .
Point the cursor to [Component] with
and then press the
key to reverse
out the cell.
Select the [H
2
O]
with
and
then
press .
Point the cursor to [Cell Length] with
and then press the
key to reverse
out the cell. Input the length of measurement
cell
with
and
then
press .
ENT
ENT
ENT