To perform an equal value calibration, you need the latest data for the load cell that you will be using. If
your load cell is not new, its actual data characteristics may be different from the initial data
characteristics that it had when it was delivered from the factory, depending on its condition and how
long it has been used. Fresh data is required to calibrate loads correctly.
③
Mode: 03
Each time you press the
key, the "Lower-limit display" will display the next user settable data name
from the following list:
"
→
→
→
→
"
Call up the data name that you want to specify data for, and then enter or change the settings.
(
See pages
20 and 23.
)
Then, turn to the "Change and select each set value" selection from the following pages. These pages
describe the operating procedures for the items below.
●
Prohibiting changes
●
Changing and storing upper and lower limit comparison values
●
Method for selecting a pair of upper and lower limits
●
Setting the power source frequency
(
IF THE UNIT WILL BE USED WITH A 60 HZ POWER
SUPPLY YOU MUST READ THIS SECTION.
)
●
Method for selecting the measurement mode
CAUTION
If your power company supplies you with 60 Hz power, make sure to change
power frequency setting from 50 to 60. The device is set to 50 Hz at delivery.