Changing and storing upper and lower comparison values
You can change the upper and lower values that are used for comparing with
the measured load. Up to eight pairs of upper and lower values can be stored
in memory. The values in the table below are already stored as default
values. (On the CM-5-003 model, a decimal point will be displayed next to the
10
1
digit.)
●
Table of default upper and lower limit values
●
Pair number
Upper value
Lower value
1
1000
100
2
1100
200
3
1200
300
4
1300
400
5
1400
500
6
1500
600
7
1600
700
8
2000
1000
1. Press and hold the
and
keys for approximately one second. (Press the
key, first.)
--- The
system will be in the pair number selection mode.
Digital load display
Lower-limit display
" " means "pair."
←
Pair number display
…
This display shows you that
Pair 1
is specified.
To change which pair is selected, press the
key. The pair number will change in the following order:
"
1→2→3→4→5→6→7→8→1
". Select the pair that you want to change.
2. Press the
key.
--- The display will change to show the upper-limit.
Digital load display
Lower-limit display
←
"
" means that you can now
change the upper-limit value.
←
Pair number display
The digit next to a flashing decimal point
can be changed.
Now for practice, try to change the current upper-limit value for pair 1.
Change the current value "1000" to "2250."
※
Depending on the model, the position of the decimal point will vary. It will either be next to the 10
1
or
the 10
2
digit. See the section, "Digital load display," on page 27.
3. First, change the value of the 10
3
digit.
Each time you press the
key, the number shown in the 10
3
digit place
(
next to the flashing decimal
point
)
will change, in the following order: "
2→3→0→1→2
".
(
The maximum number available in
the 10
3
digit place is "3."
)
Change this digit to " "