17.
Re-tighten the screws in the additional clamping system with a hex
key.
DANGER!
The additional clamping system may be loosened
due to the clamping of the Safe-Lock using the
clamping lever. Check the screw connections of
the additional clamping system and the Safe-Lock
tensioning device for tight fit before starting
grinding. Failure to observe this instruction may
lead to serious material damage or personal
injury.
DANGER!
After several minutes of grinding and in the cooling-
down phases, check that the tensioning device is firmly
connected.
If the Safe‐Lock system is correctly tensioned, the
clamping lever is automatically pulled downwards by
the eccentric when horizontal.
7.2 Grinding in inert gas with gassing lid
NOTICE!
Observe imbalance offsetting! (See
NOTICE!
Before every grinding operation, check the valves of the
gassing lid ensuring that they are clean and properly
secured. See chapter
Accessories
- 49 -