4.6 Settings in setup mode
4.6.1 Setting the imbalance check
CAUTION!
The imbalance check can be disabled at your own risk.
Fritsch assumes no liability for damage resulting from
the deactivation of the imbalance check!
NOTICE!
The highest level of sensitivity is set for the imbalance
check as the factory default. In order to adjust the
imbalance check to suit your applications as best as
possible, set the value for the imbalance check as
explained in Chapter
imbalance check“ on page 24. Please note that you
have to redefine this setting for various parameters
with different grinding sets and balls.
Activation / deactivation of the imbalance check in setup mode:
1.
Press and hold the STOP button on the front of the control panel.
2.
Switch on the device with the main switch (10) on the back of the
device. There is a brief display check. After roughly 3 seconds
setup mode is started and the STOP button can be released.
3.
If POWER SUPPLY is flashing, the device is in setup mode. If
POWER SUPPLY is not flashing, repeat the procedure.
Installation
- 24 -