FP S-Compact BB / LAN
26
Loosen the fastening on the left side (OPEN 2) in the same
way.
Take off the protective cover.
The protective cover is fastened to the device with a rubber
ring so that it cannot be lost.
Connecting the mains cable
Connection
diagram (extract)
Malfunction due to incorrect power supply wiring!
Only use conductors that meet the following specifications:
– Max. conductor cross-section size: 2,5 mm² (AWG 14)
– Copper (CU) conductors only.
Only use cables with sufficient conductor cross-section for
connection.
Damage to the connection terminals due to excessive
torque!
Use a torque screwdriver.
Tighten the clamping screws to a torque of 0.5 Nm (0.37 ft lb).
Damage to the device due to incorrect voltage connection!
Make sure that the cable is connected correctly.
Make sure that the characteristic values U = 100 - 240 V are
observed.
Connection terminals
(power connection)
Connection for
protective earth conductor