Service Manual 1001 5015 / Rev. 4
5:49
5.3.14 Checking movement of the Variable Cell
1.
Use a Multi meter tool and check between pins 6 (brown) and pin 9 (grey). The
reading should be low (close to 0 V).
2.
Issue a command to move the Cell via the Service Menu; Manual Tests /Meas-
uring Unit/ Position Settings. The reading should flicker between low (about 0
V) and high (about 5 V). When the Cell has stopped the reading should be low
again. This test indicates if the pulses are sent out to the motor.
Designation of Variable Cell Motor Pin Out
1
Red
Step motor
2
Green
Stepper motor
3
White
Stepper motor
4
Yellow
Stepper motor
5
Not used
Not used
6
Brown
GND
7
Pink
LEDs in encoder
8
Not used
Not used
9
Grey
Step
10
Blue
Home
Summary of Contents for Infratec 1241
Page 8: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 ...
Page 34: ...Infratec 1241 Grain Analyzer 2 24 Service Manual 1001 5015 Rev 4 ...
Page 68: ...Infratec 1241 Grain Analyzer 3 34 Service Manual 1001 5015 Rev 4 Fig 3 66 ...
Page 203: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 5 75 Fig 5 77 Fig 5 78 Fig 5 79 ...
Page 206: ...Infratec 1241 Grain Analyzer 5 78 Service Manual 1001 5015 Rev 4 ...
Page 208: ...Infratec 1241 Grain Analyzer 6 2 Service Manual 1001 5015 Rev 4 ...
Page 209: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 7 1 7 Schematics N A ...
Page 210: ...Infratec 1241 Grain Analyzer 7 2 Service Manual 1001 5015 Rev 4 ...
Page 212: ...Infratec 1241 Grain Analyzer 8 2 Service Manual 1001 5015 Rev 4 ...