![Foss Infratec 1241 Service Manual Download Page 170](http://html1.mh-extra.com/html/foss/infratec-1241/infratec-1241_service-manual_2323170170.webp)
5:42
Service Manual 1001 5015 / Rev. 4
5.3.7
Replacement of Connection PCB Complete
Note: The replacement is carried out with the Measuring Unit
standing apart.
1.
Disconnect the Measuring Unit Card cable from the Measuring Unit Board.
2.
Remove the Measuring Unit Card by removing the two screws holding the
Measuring Unit Card, see fig below.
Fig. 5:43 MU Card
3.
Exchange the Measuring Unit Card and assemble in reverse order.
RS
12
V
DC
1241113c