Service Manual 1001 5015 / Rev. 4
5:17
Mounting Power Supply Board
6.
Mount the new Power Supply Board into pre-scored groove on the Cooling
Flange.
7.
Insert right screw
5
and washer slightly for prepositioning purpose.
8.
Insert left screw
6
with earth cable shoe and washer for prepositioning.
Fig. 5:19
9.
Mount the Spring Plate
4
over the transistors and to the main cooling flange with
the screws
1
to
3
.
10. Tighten screws
5
and
6
.
11. Reconnect cables and refit the Power Supply accordingly.