![Foss Infratec 1241 Service Manual Download Page 189](http://html1.mh-extra.com/html/foss/infratec-1241/infratec-1241_service-manual_2323170189.webp)
Service Manual 1001 5015 / Rev. 4
5:61
5.4.5
Replacement of Locking Guide
Test Weight Control Kit P/N 1001 1779
Additional tools:
•
2 mm allen key
•
2.5 mm allen key
1.
Remove the drawer.
2.
Remove the old locking guide together with the screw
1
using a 2.5 mm allen
key.
Fig. 5:61 Removing old locking guide
3.
Log into the service menu. Open the bottom shutter.
4.
Apply locking compound on the new screw. Fit the new locking guide using a 2
mm allen key. Do not tighten the screw yet.
5.
Close the bottom shutter from the service menu.
1
Screw