5.4.3
Replacement of TWM PCB
Test Weight Control Kit P/N 1001 1779
Warning
Disconnect the incoming mains supply before removing any
cover.
NOTE! Use electrostatic discharge (ESD) protection when
handling circuit boards.
Separating the Infratec and the Test Weight Module
1.
On the rear of the instrument, remove the top screw from the plate connecting
the Infratec and the TWM.
2.
Open the right hand door and disconnect the TWM cable and drawer sensor
cable. Remove the screw inside the compartment holding the TWM and the
guiding bracket, see Fig. 5:55.
Fig. 5:55
3.
Tilt the instrument to the left onto a blanket or another soft cloth.
4.
Remove the screws holding the Cover Plate and the screws holding the TWM,
see Fig. 5:56 and Fig. 5:57.
Summary of Contents for Infratec 1241
Page 8: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 ...
Page 34: ...Infratec 1241 Grain Analyzer 2 24 Service Manual 1001 5015 Rev 4 ...
Page 68: ...Infratec 1241 Grain Analyzer 3 34 Service Manual 1001 5015 Rev 4 Fig 3 66 ...
Page 203: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 5 75 Fig 5 77 Fig 5 78 Fig 5 79 ...
Page 206: ...Infratec 1241 Grain Analyzer 5 78 Service Manual 1001 5015 Rev 4 ...
Page 208: ...Infratec 1241 Grain Analyzer 6 2 Service Manual 1001 5015 Rev 4 ...
Page 209: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 7 1 7 Schematics N A ...
Page 210: ...Infratec 1241 Grain Analyzer 7 2 Service Manual 1001 5015 Rev 4 ...
Page 212: ...Infratec 1241 Grain Analyzer 8 2 Service Manual 1001 5015 Rev 4 ...