13. Prepare the Test Weight Unit by mounting the two plastic slides from the
Infratec.
Fig. 3:29
14. Transfer the drawer sensor from the holder that you have taken out from the
Infratec (removed in step 7) to the holder for TWM (p/n 60039395). Put the
sensor as much to the left as possible. Also put the rubber sealing (p/n 15420126)
onto the left wing (see arrow below).
Fig. 3:30