5:74
Service Manual 1001 5015 / Rev. 4
5.7.2
TWM Balance Calibration
For the calibration you need the Test Weight Control Kit, p/n 10011779.
The content of the kit is:
•
2 pcs of crowned 200 g weight
•
1 pcs weighing bridge with fastening nut.
Fig. 5:75
1.
Log in on the Service menu. The password is
3677
. A hint to remember this
password is to type "FOSS" on you mobile phones keyboard.
2.
Select Manual Tests/ Test weights/ Calibration/ Mass calibration (3.6.2.2).
3.
Basically just follow the instructions given on the display.
Fig. 5:76