Service Manual 1001 5015 / Rev. 4
4:17
Line description including typical value
Units Min.
Max. Comments
SelfTest/1241xxxx/Detector/gain8/Min=Number, 495
bits
5
One
measur
ement (wi
th
clos
ed
sh
utt
er a
n
d
lam
p
tu
rn
ed
of
f)
p
er g
ain
step
is perf
or
med
, wher
e 50
00
p
o
ints ar
e r
e
ad
.
The criteria
for
this te
st
is th
at the mean value in a
ll step
s
sh
ould
b
e
b
etween 50
-2
000
.
M
eas
ur
em
ents are made in 15 steps
.
Th
e stan
dard
d
eviation
sh
ould
b
e
m
a
x
.(
see high
lighted valu
es
).
SelfTest/1241xxxx/Detector/gain8/Max=Number, 505
bits
SelfTest/1241xxxx/Detector/gain8/Mean=Number, 500
bits
50
2000
SelfTest/1241xxxx/Detector/gain8/StdDev=Number,
0.8220
bits
3.0
SelfTest/1241xxxx/Detector/gain8/Temp=Number, 24
real
0
60
SelfTest/1241xxxx/Detector/gain9/Min=Number, 495
bits
5
SelfTest/1241xxxx/Detector/gain9/Max=Number, 505
bits
SelfTest/1241xxxx/Detector/gain9/Mean=Number, 500
bits
50
2000
SelfTest/1241xxxx/Detector/gain9/StdDev=Number,
0.8220
bits
3.0
SelfTest/1241xxxx/Detector/gain9/Temp=Number, 24
real
0
60
SelfTest/1241xxxx/Detector/gain10/Min=Number, 495
bits
5
SelfTest/1241xxxx/Detector/gain10/Max=Number, 505
bits
SelfTest/1241xxxx/Detector/gain10/Mean=Number, 500
bits
50
2000
SelfTest/1241xxxx/Detector/gain10/StdDev=Number,
0.8848
bits
4.0
SelfTest/1241xxxx/Detector/gain10/Temp=Number, 24
real
0
60
SelfTest/1241xxxx/Detector/gain11/Min=Number, 492
bits
5
SelfTest/1241xxxx/Detector/gain11/Max=Number, 508
bits
SelfTest/1241xxxx/Detector/gain11/Mean=Number, 500
bits
50
2000
SelfTest/1241xxxx/Detector/gain11/StdDev=Number,
1.0310
bits
5.0
SelfTest/1241xxxx/Detector/gain11/Temp=Number, 24
real
0
60
SelfTest/1241xxxx/Detector/gain12/Min=Number, 490
bits
5
SelfTest/1241xxxx/Detector/gain12/Max=Number, 525
bits
SelfTest/1241xxxx/Detector/gain12/Mean=Number, 510
bits
50
2000
SelfTest/1241xxxx/Detector/gain12/StdDev=Number,
3.0721
bits
9.0
SelfTest/1241xxxx/Detector/gain12/Temp=Number, 24
real
0
60
SelfTest/1241xxxx/Detector/gain13/Min=Number, 525
bits
5
SelfTest/1241xxxx/Detector/gain13/Max=Number, 650
bits
SelfTest/1241xxxx/Detector/gain13/Mean=Number, 575
bits
50
2000
SelfTest/1241xxxx/Detector/gain13/StdDev=Number,
6.1721
bits
17.0
SelfTest/1241xxxx/Detector/gain13/Temp=Number, 24
real
0
60
SelfTest/1241xxxx/Detector/gain14/Min=Number, 525
bits
5
SelfTest/1241xxxx/Detector/gain14/Max=Number, 750
bits
SelfTest/1241xxxx/Detector/gain14/Mean=Number, 650
bits
50
2000
SelfTest/1241xxxx/Detector/gain14/StdDev=Number,
18.6891
bits
30.0
SelfTest/1241xxxx/Detector/gain14/Temp=Number, 24
real
0
60
SelfTest/1241xxxx/Detector/gain15/Min=Number, 500
bits
5
SelfTest/1241xxxx/Detector/gain15/Max=Number, 1000
bits
SelfTest/1241xxxx/Detector/gain15/Mean=Number, 700
bits
50
2000
SelfTest/1241xxxx/Detector/gain15/StdDev=Number,
25.6891
bits
60.0
SelfTest/1241xxxx/Detector/gain15/Temp=Number, 24
real
0
60
Summary of Contents for Infratec 1241
Page 8: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 ...
Page 34: ...Infratec 1241 Grain Analyzer 2 24 Service Manual 1001 5015 Rev 4 ...
Page 68: ...Infratec 1241 Grain Analyzer 3 34 Service Manual 1001 5015 Rev 4 Fig 3 66 ...
Page 203: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 5 75 Fig 5 77 Fig 5 78 Fig 5 79 ...
Page 206: ...Infratec 1241 Grain Analyzer 5 78 Service Manual 1001 5015 Rev 4 ...
Page 208: ...Infratec 1241 Grain Analyzer 6 2 Service Manual 1001 5015 Rev 4 ...
Page 209: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 7 1 7 Schematics N A ...
Page 210: ...Infratec 1241 Grain Analyzer 7 2 Service Manual 1001 5015 Rev 4 ...