![Foss Infratec 1241 Service Manual Download Page 30](http://html1.mh-extra.com/html/foss/infratec-1241/infratec-1241_service-manual_2323170030.webp)
2:20
Service Manual 1001 5015 / Rev. 4
Pin description
Item
Pin
Name
Description
I2C/
Power
1
SDA
I²C Serial Data
2
SCL
I²C Clock
3
GND
4
5V
5
12V
Sensor
1
12V
12V supply to external sensor
2
Trigger IN
Trigger input used to start analysis
3
GND
GND to external sensor
DSUB15
1
IO0 - Busy
Indicates that the Infratec 1241/1256 is busy
analysing
2
IO1 - Connected
Indicates that the Infratec 1241/1256 is con-
nected
3
IO7 - Trigger
Trigger input used to start analysis
4
IO6 - Reserved
Not used
5
GND
6
IO2 - AM Selection 0
Bit 0 of the 4 bit nibble used for selecting the
application model
7
IO3 - AM Selection 1
Bit 1 of the 4 bit nibble used for selecting the
application model
8
IO4 - AM Selection 2
Bit 2 of the 4 bit nibble used for selecting the
application model
9
IO5 - AM Selection 3
Bit 3 of the 4 bit nibble used for selecting the
application model
10
GND
GND to external sensor
11
12V Output
12V supply to external sensor
12
NC
NC
13
NC
NC
14
NC
NC
15
10-24V Input
External power 10-24V for opto-couplers
Summary of Contents for Infratec 1241
Page 8: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 ...
Page 34: ...Infratec 1241 Grain Analyzer 2 24 Service Manual 1001 5015 Rev 4 ...
Page 68: ...Infratec 1241 Grain Analyzer 3 34 Service Manual 1001 5015 Rev 4 Fig 3 66 ...
Page 203: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 5 75 Fig 5 77 Fig 5 78 Fig 5 79 ...
Page 206: ...Infratec 1241 Grain Analyzer 5 78 Service Manual 1001 5015 Rev 4 ...
Page 208: ...Infratec 1241 Grain Analyzer 6 2 Service Manual 1001 5015 Rev 4 ...
Page 209: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 7 1 7 Schematics N A ...
Page 210: ...Infratec 1241 Grain Analyzer 7 2 Service Manual 1001 5015 Rev 4 ...
Page 212: ...Infratec 1241 Grain Analyzer 8 2 Service Manual 1001 5015 Rev 4 ...