0x2001sub1 Status1
Table 38: Bits in 0x2001sub1 Status1
Bit
Name
Description
0
Operational
All sensors and the measuring device are ready for
operation.
1
Initializing
The bit is set when the measuring device is
initialized. The measuring device is not ready for
operation during initialization.
2
Rebooting soon
The measuring device will reboot soon.
3
Self Diagnosis
The measuring device tries to fix an error
automatically. Error messages from the
measuring device are not meaningful in this state.
4
Webserver Error
The web server is not working. The web interface
is not available.
5
Reserved
Bit reserved for future assignment.
6
Reserved
Bit reserved for future assignment.
7
Light Source Error
A light source is not working.
8
Temperature Error
At least one component outside of its permissible
operating temperature.
9
Signal Low
The signal from at least one sensor is low.
Measurements are still possible.
10
Warning
At least one warning is pending for the Blackbird
OECD.
11
Temperature
Warning
At least one component is at the limit of its
permissible operating temperature.
12 … 15
Sync counter
Counter bit.
The counter bit is incremented at each
Sync
frame.