3. Product Description
Property
Value
Unit
External sample synchrony (ETH)
(Precision Time Protocol)
< 300
μ
s
External sample synchrony (CAN)
4
< 1
ms
1
Standard deviation at a sampling frequency of 5 Hz.
2 At wavelength 1552 nm and signal level 60 %.
3 Time between power-up and operation within specification
4 Triggered by CANopen SYNC messages.
Requirements for sensors
Table 4: Requirements for sensors
Property
Value
Unit
Sensor type
Fiber Bragg grating
-
Reflectivity
70
±
20
%
Full width half maximum (FWHM)
400 ... 600
pm
Side lobe suppression
> 12
dB
Compatible sensors from fos4X
Table 5: Compatible sensors from fos4X
Measurand
Sensor
Strain
fos4Strain
fos4Strain expert
Vibration
fos4Acc 1D
fos4Acc 2D
fos4Acc 3D
Temperature
fos4Temp