7.4
Cleaning the control unit
Use a dry, soft cloth to remove dust from the Aquamarine. More stubborn
spots can be removed with a damp cloth.
You can wipe-disinfect the Aquamarine using any of the products that are
commonly used to disinfect medical electrical equipment, e.g.
MinuteWipes, Caviwipes.
Observe the instructions provided by the manufacturers of these
disinfectants.
MinuteWipes Fa. Alpro.
In USA: Caviwipes
™
.
NOTICE
Sterilize the light guide in a packing material suitable for sterilization and
storage to prevent scratching or chipping the light guides inthe
autoclave. Do not exceed a temperature of 140 °C (284 °F) during the
drying cycle. Do not abort the drying cycle before it has ended. Do not
try to accelerate the cool-down process by placing the ActiveTip in cold
water. This could cause the light guide to crack.
CAUTION
Store all components so that they are protected against
contamination.Sterilize again once the storage period has elapsed.
NOTICE
Please proceed carefully not to scratch and damage the foil on the touch
screen.
CAUTION
Spray disinfection may allow liquids to penetrate into the Aquamarine!
The Aquamarine may be disinfected only by wiping it. Do not spray
disinfect the Aquamarine laser unit.