Example: Parameters
Laser power (A)
In our example, 1.2 W power. By pushing on the 'Minus' or 'Plus' button
left respectively right to the parameter value you will be able to adjust the
emitted power between 0.2 W and 1.4 W in 0.1 W increments.
Time (B)
In our example, 0 seconds / continuous. The laser will be activated as
long as the finger switch is pressed. By pushing on the 'Minus' or 'Plus'
button left respectively right to the parameter value you will be able to
adjust the time between continuous / unlimited (0 seconds) or 1 to 9999
seconds.
Frequency (C)
In our example, CW (continuous wave mode). This is the modulation
frequency of the laser unit. By pushing on the 'Minus' or 'Plus' button left
respectively right to the parameter value you will be able to adapt the
frequency in order to change the laser operation mode. For more
informations about the operation modes, see chapter "Laser operation
WARNING
The preset power levels are considered to be safe for patients.
Increasing the power levels entails the risk of overheating the patient's
soft or hard tissue. Setting the power to excessively low levels may
result in reduced treatment efficacy.