6.2
List of preset indications
6.3
Further non-preset indications
Other indications or procedures, you may define individually and
medicate according to scientific publications, are:
● for incision: e.g. Abscesses, Frenotomy, Frenectomy, Gingival
troughing, laser assisted flap surgery, Vestibuloplasty, Implant
Uncovery
● for excision: e.g. Biopsy, Crown lengthening, Epulis, Hyperplasias,
Fibroma, Gingivectomy, Gingivoplastic, Operculectomy,
Papillectomy, Set hidden teeth free
Application
Used
wavelen
gth [nm]
Power
[W]
Mo-
dus
Duty
Cycle
[%]
Fre-
quenzy
[Hz]
Time
[s]
Fiber
[µm]
Help Menu
Incision/
Excision
(soft-tissue
surgery)
445
1,2
CW
320
Stretch the tissue and
use laser tip like an
scalpel to excise the
tissue.
Caution: Avoid contact
to implant and bone!
Perio. germ
reduction
445
0,6
CW
320
Irradiate the whole
pocket starting from the
deepest position using a
meandering course to
cover all contaminated
regions. Reduce power,
if pain sensations
appear.
Endo. germ
reduction
445
0,6
CW
200
Carefully insert the fiber
into the root canal, ca.
1 mm from the apex,
start laser and retract
fiber slowly in circular
motion from the canal
(1-2 mm/s). Repeat
procedure 4 times in 5-
seconds-intervals.
Caution: Do not stay at
the apex after laser
activation!
Haemostasis 445
1,2
CW
320
Stop the blood flow by
moving the tip of the
laser over the affected
tissue surface with
opened blood vessels.
Avoid bone contact
during treatment!