5.5
Error messages, warnings and instructions
Error code
Error message
Cause / Solution
-
PLEASE REDUCE POWER
The set power for this application exceeds the max. allowed
power for this user.
Please reduce the power.
150
PIN ALREADY EXISTS
The entered pin code already exists.
Define another pin code (user settings) or enter the correct
pin code (pin code screen)
160
PROBLEM WHEN READING THE
USER SETTINGS, RESTORED
DEFAULT VALUES
The internal plausibility check of the user settings failed. The
user settings will be overwritten with default values. This will
be automatically solved by the unit.
By repeated appearance contact the service technician.
260
PROBLEM WHEN READING THE
DEVICE SETTINGS, RESTORED
DEFAULT VALUES
The internal plausibility check of the device settings failed.
The device settings will be overwritten with default values.
This will be automatically solved by the unit.
300
PROGRAM NAME ALREADY
EXISTS
The entered program name already exists.
Please choose another program name.
400
NO CONNECTION TO THE HAND
DEVICE
The connection between control unit and handpiece is
disrupted or missing.
Check the handpiece connection at the control unit and
connect the handpiece cable if necessary
426
TEMPERATURE TOO HIGH,
PLEASE WAIT UNTIL THE HAND
DEVICE IS COOLED DOWN
Temperature too high. The handpiece back (cooling
element) has reached its upper limit.
Wait for laser handpiece and laser source to cool down.
432
CURRENT INVALID
Invalid current value. The current value measured by the
current sensor is out of the measuring range.
1. step: Software update
If error message remains: Exchange the handpiece.
451
NOT IMPLEMENTED CONNECTION
STATUS
Critical software error
Short-term solution: Restart of control unit and handpiece
Long-term solution: Software update
452
DRIVER COMMUNICATION ERROR Driver communication error
Short-term solution: Restart of control unit and handpiece
Long-term solution: Software update
453
PROTOCOL COMMUNICATION
ERROR
Communication protocol reports error during communication
on the control unit side.
Short-term solution: Restart of control unit and handpiece
Long-term solution: Software update