Sterile delivered single-use fiber tips
Labeling
Each tip is sterilized with gas (ethylene oxide). A label on the outer
packing of each set of 25 single-use fiber tips indicates the sterilization
procedure (see 3.3.1 symbols).
A green dot on the label of the outer packing serves as a process indicator
for a correct sterilization process (see 3.3.1 symbols).
Storage
To ensure the proper storage and therefore the sterility of the tips, the
following environmental factors have to be considered in terms of
storage:
● Protection from moisture
● Protection from pollution
● Mechanical stress
● Exposure to direct solar or UV radiation
● Exposure to temperature fluxuations
● In a closed storage system (e.g. cupboard, drawer), or
● in shelves or rooms of the room class II according to DIN 1946-4:
2008 -12
● From 15°C to 25°C (room temperature)
● Under relative humidity from 40% to 60% (dry conditions)
Only store the single-use fiber tips in the outer packaging, which serves
as the safety packaging (carton).
Use the oldest tips first according to their best-before month. This date is
labeled on each packaging tube of the tips and on the outer packaging of
the tip set. The remaining quantity should be stored in the closed outer
packaging (carton).
Do not refill an outer packaging (carton) of the single-use fiber tip sets
with new tips.
WARNING
Do not use the single-use fiber tips if there is no green dot on the label
of the outer packaging.