SigmaPace 1000
Operators Manual
4-6
coupling interval is slowly extended until the simulated ECG signal falls just
outside the SRP at which time it is sensed by the pacemaker, resetting the
escape interval, resulting in a longer pacing pulse interval.
Interactive Pacemaker/ ECG Simulation
Tests are performed simultaneously for both atrial and ventricular channels.
The response of the pacemaker to a wide range of simulated ECG signals is
demonstrated using this Analyzer test feature. Several variables are user
selectable to best simulate actual “real life” conditions.
Simulated ECG Rate
The Analyzer outputs either asystole or a normal sinus rhythm (NSR) at rates
covering the bradycardia, normal, and tachycardia range.
PR Interval
The PR interval of the above ECG waveform complex is also user-selectable.
Adjustable Threshold Level
When the measured pacemaker output current exceeds the threshold amplitude
setting (in milliamperes), the Analyzer outputs a simulated ventricular “paced”
beat. The pacemaker output current level required to effect “ventricular
capture” is selectable by the user.
Operational Modes
Three simulated operational modes are available in the Analyzer
Continuous
The user selects a rate, and the Analyzer generates a simulated ECG signal.
The threshold interaction with the pacemaker is observed.
Summary of Contents for SigmaPace 1000
Page 6: ......
Page 10: ...SigmaPace 1000 Operators Manual iv...
Page 12: ...SigmaPace 1000 Operators Manual vi...
Page 14: ...SigmaPace 1000 Operators Manual viii...
Page 16: ...SigmaPace 1000 Operators Manual 1 2...
Page 46: ...SigmaPace 1000 Operators Manual 2 2...
Page 56: ...SigmaPace 1000 Operators Manual 3 2...
Page 86: ...SigmaPace 1000 Operators Manual 3 32...
Page 88: ...SigmaPace 1000 Operators Manual 4 2...
Page 130: ...SigmaPace 1000 Operators Manual 4 44...