Transvenous Pacemaker Testing
Setup
and
Testing
4
4-19
F1
F2
F3
F4
eyr040.eps
4.
Press
F2
SELECT DRIVE<
to toggle between the Over and Under BPM
rates.
5.
Once a rate is selected use
F4
DRIVE RATE BPM
UP
or
DOWN
arrow
key to adjust that rate.
6.
Press
F1
HOLD
to advance the screen:
F1
F2
F3
F4
eyr016.eps
7.
Press
F4
XMIT DATA
to transmit data through the serial port.
8.
Press
F1
RELEASE
to release the hold and return to the
HOLD
screen.
9.
Press
ESC
to return to previous menu(s).