Transvenous Pacemaker Testing
Setup
and
Testing
4
4-37
•
Non-Capture
•
Non-Function
3.
Press
F1 NEXT
to advance to the following screen:
F1
F2
F3
F4
eyr048.eps
4.
Press
F1 PR
to adjust the PR Interval.
5.
Press
F2 THCH
to toggle between atrial and ventricular of the threshold
amplitude.
6.
Press
F3 THRESH
UP
or
DOWN
arrow key to adjust the selected atrial
or ventricular threshold amplitude.
7.
Press
F4 NSR RATE UP
or
DOWN
arrow key to adjust the NSR rate of
the simulation ECG signal.
8.
Conduct pacemaker testing, using the Analyzer to provide the ECG signal.
9.
Press
ESC
to return to previous menu(s).