SigmaPace 1000
Operators Manual
4-16
11.
Press
ESC
to return to previous menu(s).
Demand Mode and Dual-channel (AV) ECG Signal
The following test measures the demand mode pacemaker’s ability to interact
with a simulated dual-channel (AV) ECG signal.
F1
F2
F3
F4
eyr032.eps
1.
From the
SELECT TEST
menu, press
F2
DMAND MODE
.
F1
F2
F3
F4
eyr067.eps
2.
Press
F4 LOAD
UP
or
DOWN
arrow key to select the load.
3.
Press
F1
START
to advance the screen and start the test.