Transcutaneous Pacemaker Testing
Setup
and
Testing
3
3-23
F1
F2
F3
F4
eyr022.eps
10.
Press
F4
XMIT DATA
to transmit data.
11.
Press
F1 RESTART
to begin the test again.
12.
Press
ESC
to return to previous menu(s).
Pacemaker’s Ability to Filter Line Noise
The following test measures a pacemaker’s ability to filter line noise at either
50 or 60 Hz.
F1
F2
F3
F4
eyr012.eps
1.
From the
SELECT TEST menu
, press
F3
NOISE LINE
.