Introduction and Specifications
Instrument
Specifications
1
1-27
Battery Test Fixture ............................................ Includes the 9 volt battery supply and
facilitates connection of Analyzer to
the recessed battery terminals within
the Medtronic 5388 and 5348
Temporary Pacemakers.
Long Term Test
Test Configuration
................................................ Atrial Channel or Ventricular Channel
Pulse Count Range
............................................... 999,999 (maximum)
Rate
........................................................................ 2 % to 20 % (10 % default)
Amplitude
.............................................................. 2 % to 20 % (10 % default)
Test Time (max)
.................................................... 999:59:59 (hhh:mm:ss)
Maximum error Count
.......................................... 200
Test Termination
................................................... Manual or when maximum error count
is sensed.
Testloads Selectable
............................................ 200 , 500 , or 1000
Interactive Pacer ECG Simulation
Simulates demand, continuous, non-capture, and non-function patient ECG
activity.
Additional user-selectable parameters
PR Interval .......................................................... 0.05 to 0.300 Seconds
Independent Capture / threshold ........................ 1 mA to 25 mA (1 mA steps) for each
pacemaker channel.
Available Testloads (500
Ω
Default)
Atrial Channel
Selections ........................................................... 200
Ω
, 500
Ω
, or 1000
Ω
Accuracy ............................................................. ±1.0 % of selection
Power Rating ...................................................... 2 Watts
Ventricular Channel
Selections ........................................................... 200
Ω
, 500
Ω
, or 1000
Ω
Accuracy ............................................................. ±1.0 % of selection
Power Rating ...................................................... 2 Watts
Tracking .............................................................. Atrial and ventricular channel settings
are identical.
Summary of Contents for SigmaPace 1000
Page 6: ......
Page 10: ...SigmaPace 1000 Operators Manual iv...
Page 12: ...SigmaPace 1000 Operators Manual vi...
Page 14: ...SigmaPace 1000 Operators Manual viii...
Page 16: ...SigmaPace 1000 Operators Manual 1 2...
Page 46: ...SigmaPace 1000 Operators Manual 2 2...
Page 56: ...SigmaPace 1000 Operators Manual 3 2...
Page 86: ...SigmaPace 1000 Operators Manual 3 32...
Page 88: ...SigmaPace 1000 Operators Manual 4 2...
Page 130: ...SigmaPace 1000 Operators Manual 4 44...