SigmaPace 1000
Operators Manual
4-34
DC Load Test
F1
F2
F3
F4
eyr034.eps
1.
From the
SELECT TEST
menu, press
F2
DC LOAD
.
F1
F2
F3
F4
eyr083.eps
2.
Make necessary connections, as follows:
a.
Plug the supplied 9 V dc Load Test Cable into the matching
connector on the left side of the Analyzer.
b.
A 9 V dc Load Test Cable (Fluke Biomedical Part# 2201074) is
supplied as a standard accessory with the Analyzer. It is designed
specifically for the testing of Medtronic models 5388 Dual Channel
and 5348 Single Channel transvenous pacemakers.