SigmaPace 1000
Operators Manual
3-10
Non-Function
The Analyzer operates in “continuous mode” without “ventricular capture”
events occurring. No normal sinus rhythm (NSR) output is available in this
operational mode.
Setup and Testing
The following describes how to set up and test a transcutaneous pacemaker.
1.
Start from the
Select external pacemaker type
screen.
F1
F2
F3
F4
eyr009.eps
2.
Press
F1
NONINV
to begin Non-Invasive (transcutaneous) pacemaker
testing.