Transvenous Pacemaker Testing
Setup
and
Testing
4
4-35
c.
Remove the battery from the pacemaker and slide the black block
(with the two banana plugs) into the empty battery compartment.
d.
Apply light pressure to the pacemaker’s battery compartment door to
seat the test fixture to the recessed battery terminals.
e.
To test other brands and/or models of transvenous pacemakers, just
attach test leads (not supplied) between the test fixture banana plugs
and the exposed battery contacts.
3.
Press
F1
NEXT
to advance the screen and start the test.
F1
F2
F3
F4
eyr047.eps
4.
Press
F1
HOLD
to advance the screen:
F1
F2
F3
F4
eyr016.eps