Transcutaneous Pacemaker Testing
Setup
and
Testing
3
3-17
F1
F2
F3
F4
eyr018.eps
4.
Press
F2
SELECT DRIVE<
to toggle between the Overdrive and
Underdrive BPM rates.
5.
Once a rate is selected, press
F4
DRIVE RATE BPM
UP
or
DOWN
arrow key to adjust that rate.
6.
Press
ESC
to return to previous menu(s).
Demand Mode Pacemaker’s Ability to Sense ECG
Activity
The two following, related tests, PRP and SRP, determine the demand mode
pacemaker’s ability to sense ECG activity immediately following either a
paced event (PRP) or a sensed ECG event (SRP).
Note
A transcutaneous pacemaker must be set to Demand operation for
these tests.