![Fluke 19xC User Manual Download Page 85](http://html1.mh-extra.com/html/fluke/19xc/19xc_user-manual_2309535085.webp)
Using The Bushealth Function
Input Connections and Tested Signals
6
75
Foundation H1 Bus
Default probe setting 10:1. Use the Fluke 10:1 probe.
1
Connect the red probe to test tool input A
2
Connect the probe ground lead to the H1 bus +
3
Connect the probe tip to the H1 bus -
DATA +
DATA -
Figure 41. Foundation H1 Bus Measurement
Connections
Table 7. Foundation H1 Bus Tested Signal Properties
Signal Description
V-Level Bias
Bias voltage level
V-Level Peak-peak
voltage
Data
Bit
width
Rise
Rise time as % of bit width
Fall
Fall time as % of bit width
Jitter Jitter
distortion
Signal Dist.
Signal distortion
Noise-HF
High Frequency noise >39.1 kHz
Noise
Mid Frequency noise 7.8 ….39.1 kHz
Noise-LF
Low Frequency noise < 7.8 kHz
Note
The bus normally has continuous data traffic.
Summary of Contents for 19xC
Page 2: ......
Page 9: ...Contents continued v A Bushealth Measurements A 1...
Page 10: ......
Page 36: ...Fluke 19xC 2x5C Users Manual 26...
Page 60: ...Fluke 19xC 2x5C Users Manual 50...
Page 110: ...Fluke 19xC 2x5 Users Manual 100...
Page 146: ...Appendices Appendix Title Page Bushealth Measurements A 1...
Page 147: ......