Maintaining the Test Tool
Calibrating the Voltage Probes
9
105
Repeat steps 2 and 3 and proceed as follows:
5
Select
Probe Cal
with the arrow
keys, then accept.
A message appears asking you whether to start the 10:1
probe calibration.
6
Start the probe calibration.
A message appears telling you how to connect the probe.
Connect the red 10:1 voltage probe from the red input A
jack to the red banana jack. Connect the reference lead to
the black banana jack. (See Figure 60.)
7
Adjust the trimmer screw in the
probe housing until a pure square
wave is displayed.
8
Continue with DC calibration.
Automatic DC calibration is only
possible for 10:1 voltage probes.
The test tool automatically calibrates itself to the probe.
During calibration you should not touch the probe. A
message indicates when the DC calibration has completed
successfully.
9
Return.
Repeat the procedure for the gray 10:1 voltage probe.
Connect the grey 10:1 voltage probe from the grey input B
jack to the red banana jack. Connect the reference lead to
the black banana jack.
Note
When using 100:1 voltage probes, choose 100:1
attenuation to perform an adjustment.
Summary of Contents for 19xC
Page 2: ......
Page 9: ...Contents continued v A Bushealth Measurements A 1...
Page 10: ......
Page 36: ...Fluke 19xC 2x5C Users Manual 26...
Page 60: ...Fluke 19xC 2x5C Users Manual 50...
Page 110: ...Fluke 19xC 2x5 Users Manual 100...
Page 146: ...Appendices Appendix Title Page Bushealth Measurements A 1...
Page 147: ......