Specifications
10:1 Probe
10
127
VOLTAGE (Vrms)
FREQUENCY (kHz)
30
Figure 63. Safe Handling: Max. Voltage Between
Scope References, Between Scope References and
Meter Reference, and Between Scope
References/Meter Reference and earth ground.
10:1 Probe
Safety
Max. Input Voltage
............................ 1000 V CAT II
600 V CAT III
Max. Floating Voltage
from any terminal to earth ground............ 1000 V CAT II
600 V CAT III
Electrical specifications
Input Impedance at probe tip ..... 10 M
Ω
±
2 %//14 pF
±
2 pF
Capacity Adjustment Range........................... 10 to 22 pF
Attenuation at DC (1 M
Ω
input) ..................................10 x
Bandwidth (with FLUKE 199C)..... DC to 200 MHz (-3 dB)
Accuracy
Probe accuracy when adjusted on the test tool:
DC to 20 kHz .........................................................
±
1 %
20 kHz to 1 MHz....................................................
±
2 %
1 MHz to 25 MHz...................................................
±
3 %
For higher frequencies the probe’s roll off starts affecting
the accuracy