![Fluke 19xC User Manual Download Page 84](http://html1.mh-extra.com/html/fluke/19xc/19xc_user-manual_2309535084.webp)
Fluke 19xC-2x5C
Users Manual
74
RS-485 Bus & MOD Bus IEA-485/RS-485
Default probe setting 10:1. Use the Fluke 10:1 probes.
1
Connect the red probe to the test tool input A
and the grey probe to input B.
2
Connect the ground lead of the input A probe to
the RS-485 bus RxD/TxD N (-)
3
Connect the ground lead of the input B probe to
the RS-485 bus cable shield.
4
Connect the probe tip of both probes to the
RS-485 bus RxD/TxD P (+)
DATA +
DATA -
DATA GND
Figure 40. RS-485 Bus Measurement Connections
Table 6. RS-485 Bus Tested Signal Properties
Signal Description
V-Offset
High level voltage
V-Level Peak-peak
voltage
Data
Bit
width
Rise
Rise time as % of bit width
Fall
Fall time as % of bit width
Jitter Jitter
distortion
Signal Dist.
Overshoot
Signal distortion (Manchester
decoding, default setting)
Overshoot (NRZ decoding, can be
selected via limit setup)
Note
Continuous data traffic is not ensured . See Data
Traffic on page 70