General Specifications
Temperature
Operating ............................................................ 10
°
C to 40
°
C (50
°
F to 104
°
F)
Storage ............................................................... -20
°
C to +60
°
C (-4
°
F to +140
°
F)
Humidity
................................................................. 10 % to 90 % non-condensing
Altitude
................................................................... 3000 meters (9843 ft)
Size (L x W x H)
..................................................... 30.22 cm x 14.48 cm x 8.64 cm (11.9 in x 5.7 in x 3.4 in)
Display
................................................................... LCD Color Display
Communication
USB Device Upstream Port ................................ Mini-B connector for control by a computer
USB Host Controller Port .................................... Type A, 5 V output, 0.5 A max load. Connector for keyboard, barcode
reader, and printer
Wireless .............................................................. IEEE 802.15.4 for control by a computer
USB Device Virtual COM Port Settings
Baud Rate ....................................................... 115 200 bps
Data bits .......................................................... 8 data bits
Stop Bits ......................................................... 1 stop bit
Flow Control .................................................... Hardware (RTS/CTS)
Power
..................................................................... Lithium-Ion rechargeable, 7.2 V, 31 Wh battery, 4300 mAh
Battery Charger
..................................................... 100 V to 240 V, 50/60 Hz input, 15 V/2.0 A output. For best
performance, the battery charger should be connected to a properly
grounded ac receptacle
Battery Life
............................................................ 9 hours (minimum), 100 NIBP cycles typical
Weight
.................................................................... 1.81 kg (4 lb)
Wireless Radio
Frequency Range ............................................... 2412 MHz to 2462 MHz
Output Power ...................................................... <1 mW
Safety
..................................................................... IEC 61010-1: Overvoltage Category II, Pollution Degree 2
Electromagnetic Compatibility (EMC)
International ........................................................ IEC 61326-1: Portable Electromagnetic Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is
necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a
low-voltage power supply network that supplies buildings used for domestic purposes. There may be potential
difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated
disturbances.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test
object. The equipment may not meet the immunity requirements of this standard when test leads and/or test
probes are connected.
Korea (KCC) ....................................................... Class A Equipment (Industrial Broadcasting & Communication
Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user
should take notice of it. This equipment is intended for use in business environments and not to be used in homes.