Vital Signs Simulator
Invasive Blood Pressure Simulation and Tests
33
To set the artifact:
Note
Artifact value is not available when the Chamber value is set to
Static.
1. Push
or
to highlight the
Artifact
value on channel 1 or channel 2.
2. Push
.
3. Push
or
to highlight
5%
or
10%
if the chamber parameter is set to
Arterial
,
Radial Artery
, or
Left Ventricle
. Highlight
5 mmHg
or
10 mmHg
for all other chamber values. Highlight
Off
to turn off the artifact function.
Note
The artifact values show in percent (%) or in mmHg. These units are
set by the chamber value.
4. Push
to set the artifact value and go back to the IBP screen.
The IBP screen reappears on the display with the new artifact selection.
Simulate Invasive Blood Pressure Tests
The Product can simulate pressures that occur during a Swan-Ganz or Cardiac
Catheterization procedure. Figure 18 shows a monitor connected to the two IBP
channel jacks on the Product.
ProSim 8
Patient
Monitor
IBP Cables
glh028.eps
Figure 18. Invasive Blood Pressure Connections